version=2 dh 29 05 07 Houston 29.718 95.34101 3.23 pr 1007 inst 0 -.3849 -.3918 -.779 -2.1692 -.3 .3392 2.35 1.1364 3360 4205 3.4E-08 290 94 1713 0 5000 10000 15000 20000 25000 2972 mkiv 1 0 0 0 0 0 0 0 6100 1 242 -3 660 600 8640 178 2575 0 0 2669 250 0 64 256 0 64 38 disp 2847.55 .0744604 -5.21103E-07 2887.65 .0733695 -4.76523E-07 2924.58 .0724299 -4.45925E-07 2958.91 .0723837 -5.66067E-07 2992.67 .071959 -6.244281E-07 2814.2 .0753308 -5.47303E-07 4271.32 .111691 -7.81655E-07 4331.48 .110054 -7.14784E-07 4386.86 .108645 -6.688871E-07 4438.37 .108576 -8.49101E-07 4489 .107938 -9.36642E-07 4221.3 .112996 -8.20955E-07 zeni -.0064 -.01968 .01654 .135795 .279161 -.061978 -.464282 .452229 -.047237 co 15:41:41 user: Logged into Brewer, found unit setting at command prompt. co 15:43:15 user: BFile stopped recording at the zeni section. Will reset and restart. SWS co 15:45:39 re: Zenith discrepancy = 0 co 15:46:41 re: Azimuth discrepancy = 1 co 15:46:54 re: Found reference at step -241 Moving filterwheel #3 to o3 position at 242 co 15:59:50 ux: UV extended scan started co 16:05:06 ux: UV extended scan finished, incident DUV = 19.34832 mW/m^2 co 16:05:35 hg: Lamp not detected - resetting micrometer(s) co 16:06:06 hg: -3000 steps - QUIT - Micrometer jammed at MIN. end ? co 16:06:17 hg: Found diode at step -242 Moving filterwheel #3 to o3 position at 242 co 23:00:11 user: Loaded new ps.rtn (5-3-07) and new 07150 scheds. SWS co 23:06:41 hg: Lamp not detected - resetting micrometer(s) co 23:07:13 hg: -3000 steps - QUIT - Micrometer jammed at MIN. end ? co 23:07:24 hg: Found diode at step -242 Moving filterwheel #3 to o3 position at 242